OVERVIEW
Diameter to height ratio: 1:1~2.5; Liquid loading coefficient: 70%
Tank body: 316L stainless steel, easy to clean; stainless steel inner polishing accuracy Ra≤0.4;
Tank function: intake sterilization filter, exhaust condenser, including intake and exhaust treatment system, motor: sew AC motor for fermentation tank; double seal on upper shaft end or lower magnetic stirring to ensure zero leakage
Ventilation: Multi-channel ventilation, thermal mass flowmeter automatically controls air flow and oxygen flow
Pressure: Pressure control range: 0-0.3Mpa, pressure regulating valve adjusts the outlet pressure, digital remote display.
Stirring system: Stirring form: straight, oblique blade stirring paddle (multi-stage)
Fermentation pH control function: intelligent PID control, Mettler PH electrode;
Fermentation dissolved oxygen control function: intelligent PID control, METTLER DO electrode;
Temperature control: intelligent PID control, electric heating, water cooling, circulating pump temperature control,
Feeding: perfect feeding system (acid, alkali, foam enemy, culture agent)
Exhaust: Exhaust is filtered to inactivate emissions by heating
Fermentation broth weighing system: online detection of fermentation broth volume* (weight), load cell *Mettler (Load cell)
JIP tank cover automatic lifting system: automatic tank cover lifting system, easy to clean and maintain
LBR-SIP tank automatic sterilization system: automatic program sterilization, reducing labor intensity and improving sterilization accuracy
LBR-CIP Tank Automatic Cleaning System: Tank Automatic Cleaning System
LBR-D control system is optional: Siemens S7-200 series PLC+12-inch tablet computer touch screen display, on-site direct operation (menu type), full Chinese menu and interface; it can also transmit data with the host computer, sample and display the tank Volume data, browse analysis of system data, read settings. Control or collect parameters such as pH, DO, tank pressure, temperature, rotational speed, air flow, feed weighing, fermentation broth volume, etc., and can easily calibrate pH and DO electrodes through software; set the initial control state of the parameters; have independent Curve analysis system; direct analysis of historical curves